GF Signet 3-2725-60 DryLoc Flat ORP Electrode

The 3-2725-60 is a flat ORP electrode for oxidation-reduction potential measurement. Its sensing function differs from the pH electrodes in the adjacent 2724 and 2726 families.

A DryLoc connector and 3/4 in. NPT process thread provide the connection interfaces. Select compatible ORP sensor electronics and installation hardware for the application.

Specification
Efficiency>97% @ 25 °C (77 ° F)
Operating Range ORP±2000 mV
Compatibility2750 Electronics, 2760 Preamplifier
pH Temperature SensorPT1000 versions are compatible with Signet 2750 pH/ORP Sensor electronics for connection to a PLC or to the Signet 8900 MultiParameter Controller. , 3 KΩ Balco versions are compatible with the Signet 2760 pH/ORP preamplifier for connection to the Signet 5700 pH/ORP Monitor and the Signet 8750 pH/ORP
Process Connection¾ in. NPT , ISO 7/1-R 3/4 , Mounts into Signet fittings
Wetted MaterialspH: Ryton® (PPS), glass, UHMW PE, FPM, ORP: Ryton® (PPS), glass, UHMW PE, FPM, Platinum
Operating Temperature Range-10 °C to 85 °C (14 °F to 185 °F)
Operating Pressure Range-10 °C to 65 °C (14 °F to 149 °F): 0 to 100 psi / 65 °C to 85 °C (149 °F to 185 °F), linearity derated 100 psi to 58 psi / *Best performance for 2726-HF sensors is above 10 °C (50 °F)
Recommended Storage TemperatureThe best storage temperature for the 272X pH and ORP electrodes is 0 °C to 50 °C (32 °F to 122 °F) , The electrode glass will shatter if shipped or stored at temperature below 0 °C (32 °F) , The performance life of the electrode will shorten if stored at temperatures above 50 °C (122 °F)
In-line MountingUse the sensor threads , Use a Signet standard fitting up to 4 in. , Sensor can be mounted at any angle Submersible Mounting: , Use threads on models 2750 or 2760 , Requires ¾ inch NPT or ISO 7/1-R 3/4 male threaded liquid tight extension conduit.
Standards and ApprovalsManufactured under ISO 9001 for Quality and ISO 14001 for Environmental Management
Accessories
GF Signet 3-2700.395 pH Calibration Kit
GF Signet 3822-7115 Quinhydrone Powder for ORP Calibration
GF Signet 3-2759 pH/ORP System Tester Kit
GF Signet 3-2759.391 2751/2760 Adapter Cable
GF Signet 3-0700.390 pH Buffer Powder Kit
GF Signet 3822-7004 pH 4.01 Buffer, 473 mL
GF Signet 3822-7007 pH 7.00 Buffer, 473 mL
GF Signet 3822-7010 pH 10.00 Buffer, 473 mL